ASTM-F2118 › Historical Revision Information
Test Method for Constant Amplitude of Force Controlled Fatigue Testing of Acrylic Bone Cement Materials
The following bibliographic material is provided to assist you with your purchasing decision:
Scope
1.1 This test method describes test procedures for evaluating the constant amplitude, uniaxial, tension-compression uniform fatigue performance of acrylic bone cement materials.
1.2 This test method is relevant to orthopaedic bone cements based on acrylic resins, as specified in Specification F 451
1.3 It is not the intention of this test method to define levels of performance of these materials. Furthermore, it is not the intention of this test method to directly simulate the clinical use of these materials.
1.4 A rationale is given in Appendix X1.
1.5 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.
1.6 This standard does not purport to address all of the safety concerns associated with its use. It is the responsibility of the user of this standard to consult and establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.
Significance and Use
This test method describes a uniaxial, constant amplitude, fully reversed fatigue test to characterize the fatigue performance of a uniform cylindrical waisted specimen manufactured from acrylic bone cement.
This test method considers two approaches to evaluating the fatigue performance of bone cement:
Testing is conducted at three stress levels to characterize the general fatigue behavior of a cement over a range of stresses. The stress level and resultant cycles to failure of the specimens are plotted on an S-N diagram.
Another approach is to determine the fatigue strength of a particular cement. The fatigue strength for orthopaedic bone cement is to be determined at 5 million (5 × 106) cycles. The “two-point method” is the specified procedure for conducting fatigue testing to determine fatigue strength (1).
This test method does not define or suggest required levels of performance of bone cement. This fatigue test method is not intended to represent the clinical use of orthopaedic bone cement, but rather to characterize the material using standard and well-established methods. The user is cautioned to consider the appropriateness of this test method in view of the material being tested and its potential application.
It is widely reported that multiple clinical factors affect the fatigue performance of orthopaedic bone cement; however, the actual mechanisms involved are not well understood. Clinical factors which may affect the performance of bone cement include: temperature and humidity, mixing method, time of application, surgical technique, bone preparation, implant design, and patient factors, among others. This test method does not specifically address these clinical factors. The test method can be used to compare different acrylic bone cement formulations and products and different mixing methods and environments (that is, mixing temperature, vacuum, centrifugation, and so forth).
Keywords
acrylic bone cement; fatigue; fatigue strength; Acrylic bone cement; Constant-amplitude axial tension-compression fatigue; Fatigue; ICS Number Code 91.100.10 (Cement. Gypsum. Lime. Mortar)
To find similar documents by ASTM Volume:
13.01 (Medical and Surgical Materials and Devices)
To find similar documents by classification:
91.100.10 (Cement. Gypsum. Lime. Mortar)
This document comes with our free Notification Service, good for the life of the document.
This document is available in either Paper or PDF format.
Document Number
ASTM-F2118-03(2009)
Revision Level
2003 R09 EDITION
Status
Superseded
Modification Type
Reapproval
Publication Date
April 1, 2009
Document Type
Test Method
Page Count
8 pages
Committee Number
F04.15