Standard Practices for Calibration of Ozone Monitors Using Transfer Standards

ASTM-D5011 - 1992 R09 EDITION - SUPERSEDED
Show Complete Document History

Document Center Inc. is an authorized dealer of ASTM standards.
The following bibliographic material is provided to assist you with your purchasing decision:

Standard Practices for Calibration of Ozone Monitors Using Transfer Standards
ORDER

Scope

1.1 These practices describe means for calibrating ambient, workplace or indoor ozone monitors, using transfer standards.

1.2 These practices describe five types of transfer standards:

(A) Analytical instruments

(B) Boric acid potassium iodide (BAKI) manual analytical procedure

(C) Gas phase titration with excess nitric oxide

(D) Gas phase titration with excess ozone

(E) Ozone generator device.

1.3 These practices describe procedures to establish the authority of transfer standards: qualification, certification, and periodic recertification.

1.4 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use. See Section 8 for specific precautionary statements.

Significance and Use

The reactivity and instability of O3 precludes the storage of O3 concentration standards for any practical length of time, and precludes direct certification of O3 concentrations as SRM's. Moreover, there is no available SRM that can be readily and directly adapted to the generation of O3 standards analogous to permeation devices and standard gas cylinders for sulfur dioxide and nitrogen oxides. Dynamic generation of O3 concentrations is relatively easy with a source of ultraviolet (UV) radiation. However, accurately certifying an O3 concentration as a primary standard requires assay of the concentration by a comprehensively specified analytical procedure, which must be performed every time a standard is needed.

The primary UV standard photometers, which are usually used at a fixed location under controlled conditions, are used to certify transfer standards that are then transported to the field sites where the ambient ozone monitors are being used. See Practice D 5110.

The advantages of this procedure are:

All O3 monitors in a given network or region may be traced to a single primary standard.

The primary standard is used at only one location, under controlled conditions.

Transfer standards are more rugged and more easily portable than primary standards.

Transfer standards may be used to intercompare various primary standards.

Keywords

boric acid titration; gas phase titration; ozone; primary standard; transfer standard; Calibration--air/atmospheric analysis instruments; Ozone (O3); Photometry; Ultraviolet (UV) analysis--atmospheric applications; Ultraviolet (UV) calibration; ICS Number Code 13.040.01 (Air quality in general)

To find similar documents by ASTM Volume:

11.07 (Atmospheric Analysis)

To find similar documents by classification:

13.040.01 (Air quality in general)

This document comes with our free Notification Service, good for the life of the document.

This document is available in either Paper or PDF format.

Document Number

ASTM-D5011-92(2009)

Revision Level

1992 R09 EDITION

Status

Superseded

Modification Type

Reapproval

Publication Date

Feb. 1, 2009

Document Type

Practice

Page Count

29 pages

Committee Number

D22.03