ASTM-F2694 Standard Practice for Functional and Wear Evaluation of Motion-Preserving Lumbar Total Facet Prostheses

ASTM-F2694 - 2016 R20 EDITION - CURRENT
Show Complete Document History

Document Center Inc. is an authorized dealer of ASTM standards.
The following bibliographic material is provided to assist you with your purchasing decision:

Standard Practice for Functional and Wear Evaluation of Motion-Preserving Lumbar Total Facet Prostheses
ORDER

Price:

$63.00        


Want this as a site license?

Scope

1.1 This practice provides guidance for the functional, kinematic and wear testing of motion-preserving total facet prostheses for the lumbar spine. These implants are intended to allow motion and lend support to the functional spinal unit(s) through replacement of the natural facets.

1.2 This test method is not intended to address the bone implant interface or the static characteristics of the prosthesis components. Fatigue characteristics are included, but only as a by-product of cyclic wear testing under facet load and thus are not addressed in the typical process of generating an S-N characterization.

1.3 Biocompatibility of the materials used in a total facet prosthesis are not addressed in this practice.

1.4 The values stated in SI units are to be regarded as the standard with the exception of angular measurements, which may be reported in either degrees or radians.

1.5 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.

Significance and Use

Total Facet Prosthesis ComponentsThe total facet replacement may comprise a variety of shapes and configurations. Its forms may include, but are not limited to, ball and socket articulating joints, joints having a free-floating or semi-constrained third body, metallic load-bearing surfaces, and spring and dampening mechanisms. Additionally, it may be a unilateral or bilateral design.

Spinal Testing Apparatus:

Test ChambersIn case of a multispecimen machine, each chamber shall be isolated to prevent cross-contamination of the test specimens. The chamber shall be made entirely of corrosion resistant materials, such as acrylic plastic or stainless steel, and shall be removable from the machine for thorough cleaning between tests.

Component Clamping/FixturingSince the purpose of the test is to characterize the wear and kinematic function of the total facet prosthesis, the method for mounting components in the test chamber shall not compromise the accuracy of assessment of the weight loss or stiffness variation during the test. For example, prostheses having complicated superior and inferior surfaces for contacting bone (for example, sintered beads, hydroxylapatite (HA) coating, plasma spray) may be specially manufactured to modify that surface in a manner that does not affect the wear simulation.

The device should be securely (rigidly) attached at its bone-implant interface to the mating test fixtures.

The motion of the superior test fixture (more posterior fixture in Figs. 1 and 2) relative to the inferior testing fixture shall be constrained in three-dimensional space except for the components in the direction of specified test motions/loads.

Load and Motion:

Facet loads (fx) are initially applied in the direction of the positive X-axis.

Flexion load and motion are positive moment and rotation about the Y-axis.

Extension load and motion are negative moment and rotation about the Y-axis.

Lateral bend load and motion are positive and negative moments and rotations about the X-axis.

Axial rotation load and motion are positive and negative moments and rotations about the Z-axis.

FrequencyTest frequency is to be determined and justified by the user of this practice, and shall not exceed 2 Hz without adequate justification ensuring that the applied motion (load) profiles remain within specified tolerances and that the total facet prosthesiss wear and functional characteristics are not significantly affected. See X1.6.

Cycle CounterOne complete motion is the entire range from starting position through the range of motion (or load when in load control) and returning to the starting position (load). Cycles are to be counted using an automated counting device.


Note—This setup would require two rotational actuators and one translational actuator.

FIG. 1 Diagrams of Possible Test Apparatus for Allowing Simultaneous Lateral Bending and Axial Rotation Motions
with Anterior-Posterior Directed Facet Loading


Note—This setup would require two rotational actuators and one translational actuator.

FIG. 2 Diagrams of Possible Test Apparatus for Allowing Simultaneous Flexion-Extension and Lateral Bending Motions
with Anterior-Posterior Directed Facet Loading

Keywords

dynamic test; facet arthroplasty; posterior instrumentation; spinal implants; static test; wear assessment; weight loss method

To find similar documents by ASTM Volume:

13.02 (Emergency Medical Services; Search and Rescue; Anesthetic and Respiratory Equipment)

To find similar documents by classification:

11.040.40 (Implants for surgery, prosthetics and orthotics Including pacemakers Ophthalmic implants, see 11.040.70)

This document comes with our free Notification Service, good for the life of the document.

This document is available in either Paper or PDF format.

 

Customers who bought this document also bought:

21CFR(800-1299)
Food and Drugs

21CFR(1-99)
Food and Drugs

ISTA-2A
Performance Tests for Packaged-Products, Packaged-Products 150 lb (68 kg) or Less

Document Number

ASTM-F2694-16R20

Revision Level

2016 R20 EDITION

Status

Current

Modification Type

New

Publication Date

Nov. 17, 2020

Document Type

Practice

Page Count

9 pages

Committee Number

F04.25