ASTM-D5128 › Historical Revision Information
Standard Test Method for On-Line pH Measurement of Water of Low Conductivity
The following bibliographic material is provided to assist you with your purchasing decision:
Scope
1.1 This test method covers the precise on-line determination of pH in water samples of conductivity lower than 100 μS/cm (see Table 1 and Table 2) over the pH range of 3 to 11 (see Fig. 1), under field operating conditions. pH measurements of water of low conductivity are problematic for conventional pH electrodes, methods, and related measurement apparatus.
Sample | Sample | Sample | Δ pH Error from Additional 1 mg/L NaOH Contaminate |
0.001 | 0.055 | 7.05 | Δ 2.35 |
0.010 | 0.082 | 7.45 | Δ 1.95 |
0.100 | 0.625 | 8.40 | Δ 1.03 |
1.0 | 6.229 | 9.40 | Δ 0.30 |
8.0 | 49.830 | 10.30 | Δ 0.05 |
Sample | Sample | Sample | Δ pH Error from Additional 1 mg/L HCl Contaminate |
0.001 | 0.060 | 6.94 | Δ2.38 |
0.010 | 0.134 | 6.51 | Δ 1.95 |
0.100 | 1.166 | 5.56 | Δ 1.03 |
1.0 | 11.645 | 4.56 | Δ 0.30 |
8.0 | 93.163 | 3.66 | Δ 0.05 |
1.2 This test method includes the procedures and equipment required for the continuous pH measurement of low conductivity water sample streams including the requirements for the control of sample stream pressure, flow rate, and temperature. For off-line pH measurements in low conductivity samples, refer to Test Method D5464.
1.3 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.
1.4 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.
Significance and Use
5.1 Continuous pH measurements in low conductivity samples are sometimes required in pure water treatment using multiple pass reverse osmosis processes. Membrane rejection efficiency for several contaminants depends on pH measurement and control between passes where the conductivity is low.
5.2 Continuous pH measurement is used to monitor power plant cycle chemistry where small amounts of ammonia or amines or both are added to minimize corrosion by high temperature pure water and steam.
5.3 Conventional pH measurements are made in solutions that contain relatively large amounts of acid, base, or dissolved salts. Under these conditions, pH determinations may be made quickly and precisely. Continuous on-line pH measurements in water samples of low conductivity are more difficult (4, 5). These low ionic strength solutions are susceptible to contamination from the atmosphere, sample stream hardware, and the pH electrodes. Variations in the constituent concentration of low conductivity waters cause liquid junction potential shifts (see 3.2.1) resulting in pH measurement errors. Special precautions are required.
Keywords
high purity; low conductivity; on-line; pH; calculated pH
To find similar documents by ASTM Volume:
To find similar documents by classification:
This document comes with our free Notification Service, good for the life of the document.
This document is available in either Paper or PDF format.
Document Number
ASTM-D5128-14
Revision Level
2014 EDITION
Status
Superseded
Modification Type
Revision
Publication Date
Feb. 15, 2014
Document Type
Test Method
Page Count
11 pages
Committee Number
D19.03