Scope
1.1 This standard test method covers the measurement of the magnetically induced displacement force produced by the spatial gradients of a magnetic field on passive implants (implants that function without the supply of electrical power) and the comparison of that force to the weight of the implant.
1.2 This method does not address the issues of magnetically induced torque or RF heating.
1.3 This method is intended for devices that can be suspended from a thin string. Devices which cannot be suspended from a thin string are not covered by this test method.
1.4 The weight of the thin string from which the device is suspended during the test must be less than 1% of the weight of the tested device.
1.5 This method is applicable only to magnet systems in which the direction of the magnetic field (and the direction of the magnetically induced deflection force) is horizontal.
Keywords
mplant; metals (for surgical implants); MRI (magnetic resonance imaging); MR safety; ICS Number Code 11.040.40 (Implants for surgery, prothetics and orthotics)
To find similar documents by ASTM Volume:
13.01 (Medical and Surgical Materials and Devices)
To find similar documents by classification:
This document comes with our free Notification Service, good for the life of the document.
This document is available in either Paper or PDF format.
Document Number
ASTM-F2052-00
Revision Level
2000 EDITION
Status
Superseded
Modification Type
New
Publication Date
July 10, 2000
Document Type
Test Method
Page Count
5 pages
Committee Number
F04.15