ASTM-F2119 › Historical Revision Information
Standard Test Method for Evaluation of MR Image Artifacts from Passive Implants
The following bibliographic material is provided to assist you with your purchasing decision:
Scope
1.1 This test method characterizes the distortion and signal loss artifacts produced in a magnetic resonance (MR) image by a passive implant (implant that functions without the supply of electrical or external power). Anything not established to be MR-Safe or MR-Conditional is excluded.
Significance and Use
This test method provides a quantified measure of the image artifact produced under a standard set of scanning conditions.
This test method applies only to passive implants that have been established to be MR-Safe or MR-Conditional.
Keywords
image artifact; implant; metals for surgical implants; MRI (magnetic resonance imaging); MR compatibility; MR safety; ICS Number Code 11.040.40 (Implants for surgery, prothetics and orthotics)