ASTM-F2580 Historical Revision Information
Standard Test Method for Evaluation of Modular Connection of Proximally Fixed Femoral Hip Prosthesis

ASTM-F2580 - 2007 EDITION - SUPERSEDED
Show Complete Document History

Document Center Inc. is an authorized dealer of ASTM standards.
The following bibliographic material is provided to assist you with your purchasing decision:

Standard Practice for Evaluation of Modular Connection of Proximally Fixed Femoral Hip Prosthesis
ORDER

Scope

1.1 This test method covers a procedure for the fatigue testing of metallic femoral hip prostheses used in hip joint replacements. This test method covers the procedures for the performance of fatigue tests on metallic femoral hip stems using a cyclic, constant-amplitude force. It applies to hip prostheses that utilize proximal metaphyseal fixation and are of a modular construct, and it is intended to evaluate the fatigue performance of the modular connections in the metaphyseal filling (that is, proximal body) region of the stem.

1.2 This test method is intended to provide useful, consistent, and reproducible information about the fatigue performance of metallic hip prostheses while held in a proximally fixated manner, with the distal end not held by a potting medium.

1.3 The values stated in SI units are to be regarded as the standard. The values given in parentheses are for information only.

1.4 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.

Significance and Use

This test method can be used to describe the effects of materials, manufacturing, and design variables on the fatigue performance of metallic femoral hip prostheses subject to cyclic loading for large numbers of cycles.

The loading of femoral hip designs in vivo will, in general, differ from the loading defined in this test method. The results obtained here cannot be used to directly predict in vivo performance. However, this test method is designed to allow for comparisons between the fatigue performance of different metallic femoral hip designs, when tested under similar conditions.

In order for fatigue data on femoral hip prostheses to be comparable, reproducible, and capable of being correlated among laboratories, it is essential that uniform procedures be established.

Keywords

arthroplasty; femoral hip prostheses; orthopedic medical devices; proximal fixation; total hip arthroplasty; ICS Number Code 11.040.40 (Implants for surgery, prothetics and orthotics)

To find similar documents by ASTM Volume:

13.02 (Emergency Medical Services; Search and Rescue; Anesthetic and Respiratory Equipment)

To find similar documents by classification:

11.040.40 (Implants for surgery, prosthetics and orthotics Including pacemakers Ophthalmic implants, see 11.040.70)

This document comes with our free Notification Service, good for the life of the document.

This document is available in either Paper or PDF format.

Document Number

ASTM-F2580-07

Revision Level

2007 EDITION

Status

Superseded

Modification Type

New

Publication Date

June 15, 2007

Document Type

Test Method

Page Count

5 pages

Committee Number

F04.22