Scope
1.1 These test methods cover the determination of the radiopacity of materials and products utilizing X-ray based techniques, including fluoroscopy, angiography, CT (computed tomography) and DEXA, also known as DXA, (dual energy X-ray absorptiometry). The results of these measurements are an indication of the likelihood of locating the product within the human body.
1.2 Types of Tests - There are three methods of tests described, differing in the method of determining radiopacity.
1.2.1 Method A - Radiopacity is (1) qualitatively determined by viewing image(s) of a test sample and the image background, with or without the use of a body mimic, or ( 2) quantitatively determined as a specific difference in optical density or pixel intensity between the image of a test sample and the image background, with or without the use of a body mimic.
1.2.2 Method B - Radiopacity is determined by (1) qualitatively comparing image(s) of a test sample and a user-defined standard without the use of a body mimic, or (2) quantitatively determining the specific difference in optical density or pixel intensity between the image of a test sample and the image of a user-defined standard without the use of a body mimic.
1.2.3 Method C - Radiopacity is determined by (1) qualitatively comparing image(s) of a test sample and a user-defined standard with the use of body mimic or (2) quantitatively determining the specific difference in optical density or pixel intensity between the image of a test sample and the image of a user-defined standard with the use of a body mimic.
1.3 The values stated in SI units are to be regarded as the standard.
This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.
Significance and Use
These methods are intended to determine whether a material, product, or part of a product has the degree of radiopacity desired for its application as a medical device in the human body.
These methods allow for both qualitative and quantitative evaluation in different comparative situations.
Keywords
implants; radiopacity; surgical applications; surgical devices; surgical implants; test methods; X-ray; ICS Number Code 11.120.01 (Pharmaceutics in general); 83.080.01 (Plastics in general)
To find similar documents by ASTM Volume:
13.01 (Medical and Surgical Materials and Devices)
To find similar documents by classification:
11.120.01 (Pharmaceutics in general)
83.080.01 (Plastics in general)
This document comes with our free Notification Service, good for the life of the document.
This document is available in either Paper or PDF format.
Document Number
ASTM-F640-07
Revision Level
2007 EDITION
Status
Superseded
Modification Type
Revision
Publication Date
Nov. 1, 2007
Document Type
Test Method
Page Count
4 pages
Committee Number
F04.15