IEC-61010-2-101 › Historical Revision Information
Safety requirements for electrical equipment for measurement, control and laboratory use - Part 2-101: Particular requirements for in vitro diagnostic (IVD) medical equipment
Document Center Inc. is an authorized dealer of IEC standards.
The following bibliographic material is provided to assist you with your purchasing decision:
The following bibliographic material is provided to assist you with your purchasing decision:
IEC 61010-2-101:2015 applies to equipment intended for in vitro diagnostic (IVD) medical purposes, including self-test IVD medical purposes. It has the status of a group safety function, as specified in IEC Guide 104. This standard has been prepared in close collaboration with Working Group CENELEC BTTF 88.1. This second edition cancels and replaces the first edition published in 2002. It constitutes a technical revision and includes the following significant changes from the first edition, as well as numerous other changes:
- excluded IEC 61010-2-081 (general laboratory equipment) from the scope. This separates IEC 61010-2-081 and IEC 61010-2-101 equipment;
- updated Biohazard and Lot symbols in Table 1 in Clause 5;
- added requirement for within expiration consumables and authorized representative details in Instructions for Use to Clause 5;
- added requirement for gas or liquid markings and ratings to Clause 5;
- added requirement to include OPERATOR instructions to deal with consumable or sample spills, jams or breakage inside equipment, disposal of hazardous waste, personal protection, RISK reduction procedures relating to flammable liquids, burns from surfaces, and loading and unloading of sample and reagents in Instructions for Use to Clause 5;
- added requirement for manufacturer to provide instructions on equipment transport, storage and removal from use to Clause 5;
- added normative reference ISO 18113-5 for instructions for use of self-test IVD medical equipment in Clause 5;
- added requirement for OPERATOR maintenance instructions to Clause 7;
- added requirements for sample zones and loading zones to Clause 7;
- excluded equipment whose size and weight make unintentional movement unlikely from drop test in Clause 8;
- added requirement for biohazard marking to Clause 13;
- added requirement for interlock systems containing electric/electronic or programmable components to Clause 15;
- added informative reference to Usability standard IEC 62366 to Clause 16;
- replaced Clause 17 with requirements of ISO 14971 for RISK assessment.
- Annex BB Instructions for use for self-testing IVD Medical Equipment deleted and a reference given to ISO 18113-5 in Clause 5.
To find similar documents by classification:
This document comes with our free Notification Service, good for the life of the document.
This document is available in either Paper or PDF format.
Document Number
IEC 61010-2-101 Ed. 2.0 b:2015
Revision Level
EDITION 2.0
Status
Superseded
Publication Date
Jan. 23, 2015
Committee Number
66