ISO-21501-2 Historical Revision Information
Part 2: Light Scattering Liquid-Borne Particle Counter, Determination of Particle Size

ISO-21501-2 - 1ST EDITION - SUPERSEDED
Show Complete Document History

Document Center Inc. is an authorized dealer of ISO standards.
The following bibliographic material is provided to assist you with your purchasing decision:


ISO 21501-2:2007 describes a calibration and verification method for a light scattering liquid-borne particle counter (LSLPC), which is used to measure the size and particle number concentration of particles suspended in liquid. The light scattering method described in ISO 21501-2:2007 is based on single particle measurements. The typical size range of particles measured by this method is between 0,1 æm and 10 æm in particle size.

Instruments that conform to ISO 21501-2:2007 are used for the evaluation of the cleanliness of pure water and chemicals, as well as the measurement of number and size distribution of particles in various liquids. The measured particle size using the LSLPC depends on the refractive index of particles and medium; therefore the measured particle size is equivalent to the calibration particles in pure water.

The following are within the scope of ISO 21501-2:2007:

  • size calibration;
  • verification of size setting;
  • counting efficiency;
  • size resolution;
  • false count rate;
  • maximum particle number concentration;
  • sampling flow rate;
  • sampling time;
  • sampling volume;
  • calibration interval;
  • test report.
ORDER

To find similar documents by classification:

19.120 (Particle size analysis. Sieving Including test sieves and porosimetry)

This document comes with our free Notification Service, good for the life of the document.

This document is available in either Paper or PDF format.

Document Number

ISO 21501-2:2007

Revision Level

1ST EDITION

Status

Superseded

Publication Date

May 15, 2007

Committee Number

ISO/TC 24/SC 4